This release includes the following updates.

  • Resolved an issue where some tomographic image sets, along with their associated SR and KO files, were not viewable. In Gear View Basic 3.3.0, detection and handling of these data types have been improved. For SR images that cannot be fully read, a simplified DICOM header data image is now displayed. Blank KO images are no longer displayed. (PACS-5358)
  • Resolved an issue where series image counts could display incorrectly, or images in a study might not display. In Gear View Basic 3.3.0, multi-frame series parsing was improved, resulting in better display accuracy and performance. (PACS-534)
  • Resolved an issue where the DICOM tag for Encounter ID (0038,0010) did not populate correctly in headers when patient information was pulled from a worklist. In Gear View Basic 3.3.0, when patient demographics from a worklist are used to create a new study, the Encounter ID is correctly populated in the DICOM header. (PACS-448)